10/03/2026
🎯FAQ: Open Cup and Closed Cup Flash Point Testers-HB·ZHIWEI
1.When they test the flash point using open cup for a l**e oil where the flash point is around 250°C, how much time it takes to test the sample.
If they were to test a second sample how much time they need to wait. You can answers these questions for both closed cup and open cup
A:
When testing the open-cup flash point with an expected value of 250°C, the instrument will start heating from room temperature. The test itself takes approximately 20 minutes. After the test is completed, the instrument automatically activates the cooling fan to cool down the furnace block. The cooling time is about 20 minutes. A second experiment can be conducted once the furnace temperature has returned to room temperature.
When testing a product with a 250°C closed-cup flash point, the test for the sample takes approximately 40 minutes. Upon completion, the instrument automatically activates the cooling fan. After a cooling period of about 20 minutes, a second experiment can be conducted.
The actual time required to complete a single test cycle mainly depends on two factors:
🔸 Ambient Temperature (Affects cooling): After the test is completed, the machine needs to cool down. If your laboratory is air-conditioned (around 20°C), it cools down very quickly. However, if the room temperature is higher, approaching 30°C or above, the cooling process will naturally take a longer time.
🔸Test Procedure (Affects heating): Different procedures have different heating rates and ignition frequencies (the ignition interval might be 1°C or 2°C). Procedures with slower heating rates will naturally require more time to complete.
2.How can we do the calibration on the instruments especially. Can we do it ourselves or do they need to go to your factory?
A:Please don’t worry, you can easily do the calibration yourselves in your lab. No need to send them back to China!
You only need to use standard test oil. You run a test on the standard oil to find the temperature difference, and then simply input this difference into the machine’s software system to correct it. We will send you a step-by-step video tutorial.
You can purchase the standard flash point reference oil locally (which has a known, certified flash point value). You just need to buy the specific standard oil that corresponds to the flash point testing range of your own samples. Please rest assured, we will give you professional recommendations and provide full technical support.
---------------
📱WhatsApp: +86 18131240336
🌏Website: https://zwpowertest.com